Precaution for Use(LYZT5500 Insulation Resistance Meter)
Thank you for purchasing High Voltage
Insulation Resistance Tester. Before using the tester for the first time, to
avoid possible electric shock or personal injury, please be sure:
--- Read carefully and strictly follow the
safety regulations and precautions listed in this manual.
--- In any case, you should pay special
attention to safety when using this instrument:
·Before the test, please make sure that the
tested object is in off state or the power has been disconnected. This
instrument cannot be used for testing live equipment and devices.
·Before the test, please make sure that the
withstand voltage of the tested object must be above the output voltage gear,
and do not use the insulation resistance tester as a withstand voltage tester.
·Before the test, please make sure that the
instrument and accessories are in good condition, the instrument shell and the
insulation layer of the test wire are not damaged, exposed or broken.
·During the test, do not switch gears, and
it is strictly forbidden to touch the exposed conductor and the measuring
circuit.
·After the test, do not discharge in an
external short circuit, the tester has the function of automatic rapid discharge.
After finishing the automatic discharge, the screen shows the output voltage
lower than 36V, then the test line can be removed, to ensure that the equipment
and components are fully discharged.
·The instrument outputs high voltage, please
be sure to connect the test line first, and confirm that the connecting plug of
the test line has been tightly inserted into the instrument interface port.
Press the test button to test after the hand leaving the test line, otherwise
there is a risk of electric shock.
·Do not use the instrument when the hands
and the instrument surface are wet.
·It is strictly forbidden to use the tester
in the environment of explosive gas, steam or dust.
·Do not place and store the meter in high
temperature&humidity or dewy places and under direct sunlight for a long
time.
·Use, disassembly, calibration, and repair
of this tester must be performed by authorized personnel.
·The
instrument and the manual with the safety danger mark
, users must strictly follow the
instructions in this manual to operate safely.
1.Introduction(LYZT5500 Insulation Resistance Meter)
LYZT5500 Series High Voltage Insulation
Resistance Tester is a high performance high voltage insulation resistance
tester well researched and developed by our company. The tester has complete
and various insulation resistance index test functions, self-set test voltage
function, and good anti-interference ability. It has obvious advantage in
substation and other environments with strong induction and large distributed
capacitance. The instrument is equipped with a true-color touch screen, and all
test data and battery power are displayed on the same screen which is clear and
obvious. The combination of the rotary switch and the button makes the
operation very simple. The user does not need to memorize the operation method,
and the operation method will pop up by clicking the "HELP" button on
the screen, effectively guiding the user to operate the instrument. The tester
is suitable for insulation resistance testing of large-scale equipment such as
electrical equipment and transmission lines in large-capacity, high-voltage,
and strong-inductive environments. It is a commonly used and indispensable
instrument in the departments of electric power, telecommunications,
meteorology, computer rooms, oil fields, electromechanical installation and
maintenance, and power supply departments of industrial and mining enterprises.
The test timer of the instrument
automatically records the test time and stores the test results with date and
time. The touch screen can easily check the recorded historical data. The
completely isolated USB interface and Bluetooth communication can safely upload
the test data to the PC or mobile phone for users to analyze data. The
instrument provides the test lines of double-layer insulation and shielding
layer, and the high-voltage rod is equipped with replaceable alligator clips
and hooks, which can be adapted to various test sites. The instrument adopts a
solid double-layer shell structure, and the protection level of the case is
IP65, which can prevent the intrusion of moisture and dust during
transportation and storage, and prevent impact and effectively protect the
instrument.
2.Product Feature(LYZT5500 Insulation Resistance Meter)
2.1 5-inch touch colorful screen, the test data
is displayed on the same screen. It is very convenient to do the test operation
and view history data.
2.2 Insulation resistance range: 0.005MΩ~10.0TΩ.
2.3 Output rated voltage reaches up to 7 gears
(50V, 250V, 500V, 1kV, 2.5kV, 5kV, 10kV).
2.4 Max short circuit current is 5mA.
2.5 Large output power, strong
anti-interference ability, and stable test capability in large distributed
capacitance sites (such as long cables) and strong electromagnetic interference
environments (such as substations).
2.6 Perfect measurement feature: Insulation
resistance test (IR), polarization index test (PI), dielectric absorption ratio
test (DAR), Step test (STEP), Ramp test (RAMP), Voltage (V).
2.7 Custom setting measurement mode: PI, DAR
calculation formula can be selected to provide users with multiple choices.
2.8 Custom setting voltage mode: the test
voltage and test duration can be set freely, which is convenient for special
scenarios.
2.9 Voltage monitor: Automatically monitor the
live voltage of the measured object. Voltage exceeding 36V will automatically
prohibit the test to effectively protect the instrument and the operator.
2.10 Current monitor: Automatically displays the
current of test circuit.
2.11 Temperature monitor: Automatically displays
the temperature and humidity inside the tester.
2.12 Automatic discharge function, automatically
and quickly releases the electric charge of the tested object after the test,
with no need for an external discharge circuit.
2.13 Large capacity rechargeable lithium battery
pack 14.8V 5200mAh, to ensure long-term testing work.
2.14 Large-capacity test record memory, test
records can be browsed and analyzed directly in the tester, it can
automatically store the real-time test data with test date and test timing,
total of 1000 groups.
2.15 USB communication and upload function, the
record data can be uploaded to computer for storage and analysis.
Bluetooth communication function, which can
upload the test result to mobile phone.
Double shell structure, which is durable
and the protection level reaches IP65 (when closed).
2.16 Bluetooth communication function, which can
upload the test result to mobile phone.
2.17 Double shell structure, which is durable
and the protection level reaches IP65 (when closed).
3. Model Category(LYZT5500 Insulation Resistance Meter)
|
Model
|
Rated Voltage
|
Insulation ResistanceRange
|
Short Circuit Current
|
|
LYZT5500
|
50V, 250V, 500V, 1kV, 2.5kV, 5kV
|
0.005MΩ~5.00TΩ
|
5mA Max
|
|
LYZT5500B
|
50V, 250V, 500V, 1kV, 2.5kV, 5kV, 10kV
|
0.005MΩ~10.0TΩ
|
5mA Max
|
4. Insulation Resistance
Range & Accuracy(LYZT5500 Insulation Resistance Meter)
|
Rated
Voltage
|
Insulation Resistance Range
|
Accuracy
|
Remark
|
|
50V
|
0.005MΩ~5.00GΩ
|
±5%rdg±3dgt
|
|
|
5.00GΩ~50.0GΩ
|
±10%rdg±3dgt
|
|
|
250V
|
0.05MΩ~25.0GΩ
|
±5%rdg±3dgt
|
|
|
25.0GΩ~250GΩ
|
±10%rdg±3dgt
|
|
|
500V
|
0.10MΩ~50.0GΩ
|
±5%rdg±3dgt
|
|
|
50.0GΩ~500GΩ
|
±10%rdg±3dgt
|
|
|
1kV
|
0.50MΩ~100GΩ
|
±5%rdg±3dgt
|
|
|
100GΩ~1.00TΩ
|
±10%rdg±3dgt
|
|
|
2.5kV
|
1.00MΩ~250GΩ
|
±5%rdg±3dgt
|
|
|
250GΩ~2.50TΩ
|
±10%rdg±3dgt
|
|
|
5kV
|
2.00MΩ~500GΩ
|
±5%rdg±3dgt
|
|
|
500GΩ~5.00TΩ
|
±10%rdg±3dgt
|
|
|
10kV
|
5.00MΩ~1.00TΩ
|
±5%rdg±3dgt
|
|
|
1.00TΩ~10.0TΩ
|
±10%rdg±3dgt
|
|
5. Technical Specification(LYZT5500 Insulation Resistance Meter)
|
Function
|
Insulation resistance measurement (IR);
polarization index measurement (PI); absorption ratio (DAR); Step test
(STEP); Ramp test (RAMP); Voltage measurement (V).
|
|
Power Supply
|
Rechargeable lithium battery 14.8V 5.2Ah
|
|
Display Screen
|
5-inch touch colorful screen, display
area: 107mm × 58mm
|
|
Rated Voltage
|
LYZT5500: 50V, 250V, 500V, 1kV, 2.5kV,
5kV; LYZT5500B: 50V, 250V, 500V, 1kV, 2.5kV, 5kV, 10kV
|
|
Custom Setting Voltage
|
LYZT5500: 40V~5kV, LYZT5500B: 40V~10kV
|
|
Output Voltage Accuracy
|
+4%±10V (when 50V: +4%±2V)
|
|
Insulation Resistance Test Range
|
LYZT5500: 0.005MΩ~5.00TΩ; LYZT5500B:
0.005MΩ~10.0TΩ
|
|
Short Circuit Current
|
5mA MAX
|
|
Polarization Index (PI)
|
Yes
|
|
Dielectric Absorption Ratio (DAR)
|
Yes
|
|
Step Test (STEP)
|
Yes
|
|
Ramp Test (RAMP)
|
Yes
|
|
Voltage Test
|
Range: DC:0V~1000V; Accuracy: ±5%rdg±3V;
Range: AC: 0V~750V; Accuracy: ±5%rdg±3V
|
|
Real Time Current Display
|
Display test current value, current
display range: 0.001nA~5mA
|
|
Real Time Voltage Display
|
Real time monitoring the voltage of the
tested terminal. In the non-test state, when the voltage is larger than 36V,
the test is prohibited.
|
|
Test Timer
|
Automatic record test time, time range:
0s~9999s
|
|
Test Time
|
After selecting the DAR, PI test options,
the test will be stopped immediately after the test result is obtained; in
the custom test mode, the test duration can be set by user; in the normal
test mode, the test time is not limited.
|
|
Custom Setting Time
|
10s~3600s
|
|
Store Function
|
Automatically store the test data, which
can replay the test record in tester.
|
|
Bluetooth Printer
|
This function is optional. (The printer
can be connected in the interface for setting the clock and calculating
formulas)
|
|
Analog pointer display
|
This function is optional.
|
|
Battery Power Display
|
With battery power display, when battery
voltage is low, it will remind to charge in time.
|
|
Automatic Shutdown
|
15 minutes after startup, the tester will
shut down automatically without any operation.
|
|
Maximum immunity to noise/interference
|
6mA
|
|
Insulation Resistance
|
50MΩ (1000v) (between the test circuit
and shell)
|
|
Withstand Voltage
|
AC 3kV 50Hz 1min (between the test
circuit and shell)
|
|
USB Communication / Bluetooth
Communication
|
USB Communication: Yes, the test
record can be uploaded to the PC through USB cable. Bluetooth
Communication: Yes, can be connected to Android phones or other devices
with Bluetooth communication function.
|
|
Meter Dimensions
|
Approx. 280mm × 260mm × 160mm
|
|
Meter Weight
|
LYZT5500: Approx. 3140g (including
battery); LYZT5500B: Approx. 3180g (including battery); Total weight: Approx.
6.2kg (including accessories)
|
|
Operating Temperature & Humidity
|
-20℃~50℃;
below 80%RH
|
|
Storage Temperature & Humidity
|
-25℃~65℃;
below 80%RH
|
|
Protection Level
|
Closed case: IP65, opened case: IP40
|
|
Suitable Safety Standard
|
IEC61010-1, IEC61326-1
|
6. Tester Structure
|
1
|
Outside case
|
|
2
|
LINE high voltage port
|
|
3
|
GUARD protection port
|
|
4
|
EARTH earth port
|
|
5
|
Charge interface
|
|
6
|
USB interface
|
|
7
|
Touch colorful screen
|
|
8
|
Function selection button
|
|
9
|
Measurement range selection switch
|
|
10
|
Test button
|
|
11
|
Alligator clip test line
|
|
12
|
High voltage test rod
|
|
13
|
High voltage rod replace clip
|
|
14
|
High voltage rod replace hook
|
7. Safety Tips
To avoid electric shock or personal injury,
follow these guidelines:
7.1 Do not work alone or in explosive gas,
steam or high dust environment.
7.2 Before testing, make sure that the tester
voltage display is not higher than 36V.
7.3 Confirm that the tested device is in a
power-off state. This instrument is not allowed to be used to test live
devices.
7.4 Confirm that the withstand voltage of the
tested device is within the output voltage range of the selected test gear.
7.5 After the test, do not take away the test
line when the voltage is higher than 36V to ensure that the distributed
capacitor electric charge is completely discharged.
7.6 Check whether the insulation of the test
line is damaged, if damaged, please replace the cable.
7.7 Insert the test line into the correct port,
otherwise high voltage may be output to dangerous parts.
7.8 During the test, do not touch the test line
clip and the connected part of the tested device. Touching may cause electric
shock accident.
7.9 When removing the test line after the test,
please hold the finger protector back end of the test line clip.
8. Measurement Method
8.1. Measurement Steps
1)Non-destructive inspection of the
instrument, and confirm the tester and test line are not damaged.
2)Check the battery voltage, check the
battery power indicator after starting up the tester, if the power is too low,
please charge first.
3)Check the storage space for test records,
and check the used rate of the storage space in the lower right corner after
starting up the tester. If the remaining space is small, it should be cleared
in time.
4)Correct wiring connection, insert the test
cable into the corresponding jack, and connect the tested device.
5)Check whether the test device is live or
not, and it is strictly forbidden to test when the device is live.
6)Confirm the test voltage gear to prevent
the tested device from being damaged under the selected voltage gear.
7)Set the measurement mode, can choose the
calculation formula when measuring DAR and PI parameters.
8)Test and record the measurement results.
9)After the test, the instrument should be
cleaned of dirt and properly stored in a ventilated and dry place.
8.2. Button & Rotary Knob Function

|
Figure 2 - Knob Position Diagram
Figure 3 - Button Diagram
Normal test: press the TEST button directly
to start the test after selecting the voltage gear, and press
the TEST button again to stop the test. This is called normal test.
|
If other test modes or test items are
needed, press the corresponding button before the test, and then press the TEST
button to start the test; it is not allowed to change the test mode or option
during the test.
RAMP and STEP modes can only
be selected in preset gear, and these two modes are not allowed to be selected
in V gear and VS gear.
When selecting the dielectric absorption
ratio (DAR) or polarization index (PI), there is a test duration requirement,
and the test will be automatically stopped after the corresponding test result
is collected; if the test is stopped by pressing the test button in advance
during testing, the corresponding test option will be discarded and no test
result will be obtained.
8.3. Test Interface
1)Normal measurement mode interface
|
1
|
Present voltage gear and range
|
|
2
|
Present output terminal state
(Red-output, Yellow-voltage over 36V)
|
|
3
|
Insulation resistance IR
|
|
4
|
Dielectric absorption ratio DAR
|
|
5
|
Polarization index PI
|
|
6
|
Number of record groups (test record
usage statistics)
|
|
7
|
Estimated test duration (minute: second,
not displayed if stopped manually)
|
|
8
|
HELP button
|
|
9
|
Bluetooth, USB, battery power indication
|
|
10
|
Test time (minute: second)
|
|
11
|
Transient current value
|
|
12
|
Transient voltage value
|
|
13
|
Ambient temperature and humidity during
the test (inside housing)
|
2)Custom setting output voltage measurement
mode interface
|
1
|
Present output terminal state
(Red-output, Yellow-voltage over 36V)
|
|
2
|
Insulation resistance IR
|
|
3
|
Dielectric absorption ratio DAR
|
|
4
|
Polarization index PI
|
|
5
|
Number of record groups (test record
usage statistics)
|
|
6
|
Test time (minute: second)
|
|
7
|
HELP button
|
|
8
|
Bluetooth, USB, battery power indication
|
|
9
|
Test time setting
|
|
10
|
Transient current value
|
|
11
|
Output voltage setting
|
|
12
|
Transient Voltage Value
|
3)Step (STEP) measurement mode interface
|
1
|
Gear test range
|
|
2
|
Insulation resistance curve median value
|
|
3
|
Green curve is resistance, yellow curve
is voltage
|
|
4
|
The MAX. & MIN. value of insulation
resistance
|
|
5
|
Number of record groups (test record
usage statistics)
|
|
6
|
Check step test record
|
|
7
|
Voltage curve median value
|
|
8
|
Step progress
|
|
9
|
Step voltage
|
|
10
|
HELP button
|
|
11
|
Bluetooth, USB, battery power indication
|
|
12
|
Test time
|
|
13
|
Transient current value
|
|
14
|
Transient voltage value
|
4)Ramp (RAMP) measurement mode interface
|
1
|
Gear
test range
|
|
2
|
Insulation
resistance curve median value
|
|
3
|
Green curve is resistance, yellow curve is voltage
|
|
4
|
The MAX. & MIN. value of
insulation resistance
|
|
5
|
Number
of record groups (test record usage statistics)
|
|
6
|
Check
ramp test record
|
|
7
|
Voltage
curve median value
|
|
8
|
Step
progress
|
|
9
|
Ramp
target voltage (stop voltage)
|
|
10
|
HELP button
|
|
11
|
Bluetooth,
USB, battery power indication
|
|
12
|
Test
time
|
|
13
|
Transient
current value
|
|
14
|
Transient
voltage value
|
The principle of insulation resistance
measurement is that a voltage V is generated by the high voltage generator and
applied to both ends of the tested resistance. By measuring the current I
flowing through the resistance, the resistance value R is calculated according
to ohm's law.
R=V/I
8.4. Insulation Resistance (IR) Measurement
The principle of insulation resistance
measurement is that a voltage V is generated by the high voltage generator and
applied to both ends of the tested resistance. By measuring the current I
flowing through the resistance, the resistance value R is calculated according
to ohm's law.
8.5. Polarization Index (PI) Measurement
Polarization Index (PI) refers to the ratio
of insulation resistance between 10 minutes and 1 minute. The polarization
index test takes 10 minutes to complete. The polarization test will be
completed and saved when the insulation test is 10 minutes or longer. In the
custom setting mode, the user can select the calculation formula of PI.

|
Polarization Index (PI)
|
>4
|
4~2
|
2.0~1.0
|
<1.0
|
|
Insulation State
|
Very Good
|
Good
|
Problems
|
Bad
|
8.6. Dielectric Absorption Ratio (DAR) Measurement
Dielectric Absorption Ratio refers to the
ratio of insulation resistance between 1 minute and 15 seconds. Dielectric
absorption ratio needs to be completed within 1 minute. Therefore, for all
insulation tests less than 1 minute, the measurement data will be stored as
invalid data. When the insulation test time is 1 minute or longer, the
absorption ratio is included in the results. In the custom setting mode, the
user can select the calculation formula of DAR.
|
Dielectric Absorption Ratio(DAR)
|
>1.4
|
1.25~1.0
|
<1.0
|
|
Insulation State
|
Very Good
|
Good
|
Bad
|
8.7. Ramp(RAMP) Measurement
Ramp test mode is an automated test
process. After the test starts, the output voltage gradually rises to the
selected voltage gear, and the instrument records the insulation resistance
value every second during the voltage rising process. When the insulator is
defective, the resistance value will decrease along with the increase of the
applied voltage and time; when the insulator is not defective, the insulation
resistance will basically remain unchanged during the voltage rising process.
It should be noted that when selecting a
lower gear voltage and in the initial stage of voltage rising, the voltage
fluctuation will be large, because the high voltage output is difficult to
control in the low voltage stage, which is a normal phenomenon.
8.8. Step(SETP) Measurement
Step test mode is an automated test
process. After the test starts, the output voltage gradually rises with 10% of
the selected voltage gear, and the instrument records the insulation resistance
value every step during the voltage rising process. When the insulator is
defective, the resistance value will decrease along with the increase of the
applied voltage and time; when the insulator is not defective, the insulation
resistance will basically remain unchanged during the voltage rising process.
It should be noted that when selecting a
lower gear voltage and in the initial stage of voltage rising, the voltage
fluctuation will be large, because the high voltage output is difficult to
control in the low voltage stage, which is a normal phenomenon.
8.9. Voltage(V) Measurement
This instrument has the function of
measuring AC and DC voltage. When measuring insulation resistance, this
function is only used to help check whether the tested circuit is live. Do not
measure AC or DC voltage above 1000V to avoid damaging the tester.
When testing the voltage, turn the rotary
knob to the voltage test gear "V" to enter the voltage measurement
mode, connect the high voltage red test line to the LINE port, and connect the
ground black line to the EARTH port. As shown in Figure below:

|
1
|
AC/DC and DC ± wiring connection
indication
|
|
2
|
AC/DC voltage TRMS
|
|
3
|
AC power frequency type, automatically
judge the measured power frequency, and higher frequencies are all regarded
as DC systems.
|
|
4
|
When the initial voltage value is not 0V,
click to calibrate it to 0V.
|
|
5
|
Bluetooth, USB, battery power indication
|
|
When measuring DC voltage, connect the
red test line to the positive pole and the black test line to the negative
pole. If the connection is reversed, there will be a prompt at position 1.
Please correct it immediately. Long-term reverse connection of external DC
voltage will cause irreversible damage to the tester.
|
|
Do not measure AC/DC voltage above 1000V
to avoid damaging the tester.
|
8.10. Setting
The tester can set the date and time, and
select the calculation formula.
When modifying the date and time, slide
left on the screen to switch to the test record browsing page, then slide left
again to enter the clock setting interface. Slide up and down to select the
time, then press confirm. As shown in Figure 13.
When modifying the formula, check the
calculation formula to be used. After all the settings are modified and
finished, press the Accept button to save the modification and take effect and
return to the previous page.
Calibration requires precision instruments
and specialized skills, the content is beyond the scope of this manual.
|
1
|
Date and time setting
|
|
2
|
Accept and save the setting value
|
|
3
|
Give up the modify value and return back
to test interface
|
|
4
|
Calibration procedure, for professional
use only
|
|
5
|
PI formula select
|
|
6
|
DAR formula select
|
|
7
|
Bluetooth, USB, battery power indication
|
|
8
|
Date and time setting
|
9.Consult Test Record
In each measurement mode, slide the screen
to the left, and the screen switches to the test record browsing page.
9.1. Normal Test Mode Record Interface
|
1
|
Test
record (one set test record per line)
|
|
2
|
Bluetooth,
USB, battery power indication
|
|
3
|
Skip
to the first test record
|
|
4
|
Skip
to the previous page test record
|
|
5
|
Skip
to the next page test record
|
|
6
|
Skip
to the last test record
|
|
7
|
Delete
all test records
|
|
8
|
Test
record usage statistics
|
9.2. Ramp Test Record Interface
|
1
|
Test record summary
|
|
2
|
Insulation resistance value
|
|
3
|
Green curve is resistance, yellow curve
is voltage
|
|
4
|
The MAX. & MIN. value of insulation
resistance
|
|
5
|
Cursor move left
|
|
6
|
Cursor move right
|
|
7
|
Exit the test record browse interface
|
|
8
|
Test record usage statistics
|
|
9
|
Present cursor test data
|
|
10
|
Present record number
|
|
11
|
Delete all test records
|
|
12
|
Skip to the last test record
|
|
13
|
Skip to the next test record
|
|
14
|
Skip to the previous test record
|
|
15
|
Skip to the first test record
|
9.3. Step Test Record Interface
|
1
|
Test record summary
|
|
2
|
Insulation resistance value
|
|
3
|
Green curve is resistance, yellow curve
is voltage
|
|
4
|
The MAX. & MIN. value of insulation
resistance
|
|
5
|
Cursor move left
|
|
6
|
Cursor move right
|
|
7
|
Exit the test record browse interface
|
|
8
|
Test record usage statistics
|
|
9
|
Present cursor test data
|
|
10
|
Present record number and step voltage
|
|
11
|
Delete all test records
|
|
12
|
Skip to the last test record
|
|
13
|
Skip to the next test record
|
|
14
|
Skip to the previous test record
|
|
15
|
Skip to the first test record
|
10. Normal Wiring Connection Method
10.1.The Insulation Resistance Test
of Cable
1)Normal two-wires measurement connection
mode
There is leakage current on the surface of
the inner insulation layer near the end of the cable. This leakage current is
also in the measurement current of the "-" terminal, which will make
the measurement resistance reading value lower than the actual insulation
resistance value. Non-ultra high resistance values can be measured in this
way.
2)Ultra high resistance value 3-wire
measurement connection mode
Wrapping the bare metal wire with good
conductivity around the periphery of the inner insulation layer, and prevent
the leakage current on the surface of the measured object by connecting the
safety terminal to the outer conductor of the inner insulation layer. The
surface leakage current will be directed to the safety terminal in order to
eliminate the surface leakage current on the measurement path between the +/-
poles and improve the accuracy of the measurement reading.
3)Ultra high resistance value 3-wire
insulation resistance measurement connection mode
Wrapping the bare metal wires with good conductivity around the
periphery of the inner insulation layer and connecting safety terminals to the
outer conductors of the inner insulation and unused wire. The surface leakage
current will be directed to the safety terminal in order to eliminate the
surface leakage current on the measurement path between the +/- poles, which
ensures that the measured insulation resistance is the insulation resistance
between the selected cable and the outer insulation resistance, while eliminating
leakage paths between cables.
4)Ultra high resistance value 3-wire
insulation resistance measurement connection mode
Wrapping the bare metal wires with good
conductivity around the periphery of the inner insulation and connecting safety
terminals to the outer conductors of the inner insulation layer. The surface
leakage current will be directed to the safety terminal in order to eliminate
the surface leakage current on the measurement path between the +/- poles,
which ensures that the measured insulation resistance is the insulation
resistance between the selected cable, while eliminating leakage paths between
cables.
10.2 Transformer Insulation
Resistance Test
1)The insulation resistance test between
primary winding and secondary winding to ground
2)The insulation resistance test between the
primary winding (grounded) and the secondary winding
3)Insulation resistance test between
secondary windings
11. Bluetooth Communication
11.1. Software Introduction
This software is a mobile phone APP that
only supports mobile phones with Android system 4.0 or above. It is a Bluetooth
APP used to connect to this instrument and does not support other devices with
Bluetooth function.
11.2. Software Installation
Use the mobile browser to scan the QR code
to obtain the APP software download link. Need to obtain the corresponding
permissions during installation. Please click Agree or Allow.
11.3. Software Connection
After downloading the software, open the
software and turn on the mobile phone Bluetooth, find the device whose name
begins with HVIR, and click Connect to operate.
12. USB Communication
Before use, install the USB driver and the
data upload software of this instrument.
Log in to the official website to download
the data upload software of the corresponding model.
13. Battery And Charge
When the battery power is only 1-2 grids,
please charge in time. The charge indicator will flash green when charging, and
stay green when fully charged.
Shutdown and confirm that the tester is in
off state, and connect the charger to charge through the mains.
If the tester will not be used for a long
time, please charge every 2 months.
The standard charger must be used for
charging. After charging, please put away the charger so that it will be
convenient to use next time.
14. Instrument Maintenance
To avoid electric shock or personal injury
Do not try to repair or maintain the
instrument out of scope described in this manual.
This instrument is for maintenance by
authorized personnel only.
The user shall not replace any part of the
instrument without permission.
The accuracy of the instrument can be
maintained for 1 year after calibration at 23℃±5℃ (operating temperature: 0℃~35℃). For operating temperatures outside the range
(-20℃ to 0℃ and 35℃ to 50℃), error increases by ±0.25% per ℃.
Clean: In order to avoid the risk of
electric shock or personal injury, before cleaning the instrument, you should
wring out the cloth and do not let any terminal touch water. Clean the shell
regularly with a mild cleaner. Do not use abrasive or solvent to clean this
instrument.
Store: After use, the instrument should be
stored in a dry and clean environment.
|
Use, disassembly, calibration, and repair
of this tester must be performed by authorized personnel.
|
15. Accessories
|
Tester
|
1 PC
|
|
High Voltage Test Rod (Red, 3m)
|
1 PC
|
|
Alligator Clip Test Line (Black and
Green, 3m each)
|
2 PCs
|
|
Alligator Clip (spare)
|
1 PC
|
|
Hook
|
1 PC
|
|
USB Communication Cable
|
1 PC
|
|
Charger Adaptor (DC 19V)
|
1 PC
|
|
Instrument Bag
|
1 PC
|
|
Manual / Warranty Card / Qualification
Certificate
|
1 Set
|