1. characteristics
and performance
The automatic anti-interference different
frequency dielectric loss tester is used for on-site anti-interference
dielectric loss measurement, or precision dielectric loss measurement in the
laboratory. The instrument is an integrated structure with built-in dielectric
loss bridge, variable frequency power supply, test transformer and standard
capacitors. Using frequency conversion anti-interference and Fourier transform
digital filtering technology, automatic and intelligent measurement, the
measurement data is very stable under strong interference. The measurement
result is displayed by the large-screen LCD, and the built-in micro-printer can
be printed out。
1.1 dielectric loss tester Main technical indicators
1.1.1Dielectric
loss and capacitance measurement
Accuracy: Cx: ±(reading×1%+1pF)
tgδ: ±(reading×1%+0.00040)
Anti-interference index: Frequency conversion
anti-interference,The above accuracy can still
be achieved under 200% interference
Capacity range: Internal voltage: 3pF~60000pF/10kV 60pF~1μF/0.5kV
External voltage: 3pF~1.2μF/10kV 60pF~30μF/0.5kV
Resolution: Up to 0.001pF, 4
significant digits
Tgδ range: Unlimited, resolution of
0.001%, automatic identification of three types of samples: capacitance,
inductance and resistance。
Test current range:10μA~5A
Internal voltage: Set voltage range: 0.5~10kV
Maximum
output current: 200mA
Lifting
and pressing votlage method: continuous and smooth adjustment
Voltage
accuracy: ±(1.5%×reading+10V)
Voltage
resolution: 1V
Test frequency: 45~65Hz Integer frequency
49/51Hz、45/55Hz Automatic double
frequency conversion
Frequency
accuracy:±0.01Hz
External voltage: Maximum test current during positive
connection1A / 40~70Hz
Maximum
test current when reverse wiring10kV / 1A / 40~70Hz
CVT Self-excited low voltage output: output voltage 3~50V, output
current3~30A
Measuring time: about 30s,Related to measurement
method
1.1.2
Other indicators
Input power: 180V~270VAC,50Hz/60Hz±1%,Mains or generator power
supply
Computer interface: Standard RS232 interface (optional)
Printer: Comes with mini thermal printer
Ambient temperature: -10℃~50℃
Relative humidity: <90%, no
condensation
dielectric loss tester Brief table of main
technical indicators for selection
|
Capacity range
|
Maximum output current mA
|
Size
LxWxH
cm
|
Weight
kg
|
High voltage dielectric loss
|
CVT
Self-excited measurement
|
Reverse wiring low-voltage side
shield
|
Circuit discharge reminder
|
Printer
|
|
|
3~60k
|
200/
10kV
|
|
22
|
support
|
No need for external equipment C1/C2 to measure the high voltage
connection at the same time
|
C1/C2
Simultaneous measurement
|
have
|
Thermal
|
RS232
Store 100 groups
|
1.2
Main features of capacitance and dielectric loss measurement
1.2.1Frequency
conversion anti-interference
Using
frequency conversion anti-interference technology, it can still measure
accurately under 200% interference, and the test data is stable, suitable for
on-site anti-interference dielectric loss test。
1.2.2High-precision measurement
Using techniques
such as frequency floating, digital waveform analysis and bridge
self-calibration, combined with high-precision three-terminal standard
capacitors, high-precision dielectric loss measurement is achieved, and the
accuracy and stability of the positive/reverse wiring measurement are
consistent.。
The input
resistance of all ranges of the instrument is lower than 2Ω, which eliminates
the influence of the additional capacitance of the test line。
It can be
connected to the oil cup for precision insulation oil dielectric loss test, and
it can be connected to the solid material measuring electrode for precision
insulation material dielectric loss test。
1.2.3 Good compatibility
Automatically
identify 50Hz/60Hz system power supply, support generator power supply, even if
the frequency fluctuates greatly, it can be measured normally.
Built-in series
and parallel dielectric loss measurement models, which can be fully compatible
with the calibration table and dielectric loss standard device, which is
convenient for instrument verification。
1.2.4 Multi-level safety protection to ensure personal and
equipment safety
High-voltage
protection: short-circuit, breakdown or high-voltage current fluctuations of
the test product can cut off the output at a high speed in a short-circuit
mode.
Low
voltage protection: wrong connection of 380V, power supply fluctuation or
sudden power failure, start protection, will not cause over voltage.
Grounding
protection: When the instrument's poor grounding causes the enclosure to carry
dangerous voltage, the grounding protection is activated.
C
V T: Four protection limits of high voltage and current, low voltage and
current, which will not damage the equipment; incorrect selection of the menu
will not output excitation voltage. There is no 10kV high voltage output during
CVT measurement.
Misoperation
prevention: two-level power switch; real-time monitoring of voltage and
current; multiple keystrokes to confirm; terminal high/low voltage is clearly
defined; slow voltage increase, rapid voltage reduction, sound and light alarm.
Anti-"capacity
rise": When measuring large-capacity samples, there will be a
"capacity rise" effect of increased voltage. The instrument can
automatically track the output voltage and keep the test voltage constant.
Anti-seismic
performance: The instrument adopts a unique anti-seismic design, which can
withstand strong long-distance transportation vibration and bumps without being
damaged.
High-voltage
cable: It is a high-voltage-resistant insulated wire that can be used for
mopping the ground.
1.2.5 Technical breakthrough, powerful function
(1) With an external standard capacitor
interface, it can automatically track the frequency of the external test power
supply from 40Hz to 70Hz, and support industrial frequency power supply and
series resonance power supply for large-capacity high-voltage dielectric loss
test.
(2) It has the function of discharging prompt
for poor circuit contact to facilitate the judgment of whether the wiring is
reliable.
(3) With CVT self-excitation measurement
function, C1/C2 can be measured at the same time during one connection,
automatically compensate for the effect of busbar grounding and the partial
voltage of standard capacitors, without changing wires and connecting any
external accessories, and high-voltage connecting wires can be grounded.
(4) With reverse wiring low-voltage shielding
function, when the 220kVCVT bus is grounded, the 10kV reverse wiring dielectric
loss measurement can be performed on C11 without disconnecting the wiring, and
the master and slave capacitors can be measured at the same time for one
wiring.
(5) Chinese graphic menu, large backlit LCD
display.
(6) Equipped with thermal printer, printing
data is clear and fast, without noise.
(7) With calendar clock, 100 sets of
measurement data can be stored.
(8) With computer interface (optional).
Through this interface, measurement, data processing and report output can be
realized, and the internal measurement software of the instrument can also be
upgraded. A computer can control 32 instruments, which can be integrated into a
comprehensive high-voltage test vehicle。